Mooring Details: from 2022-10-17 16:05:00 to 2024-05-06 16:06:00
Time Event: 2022-10-17 16:05:00, Deployed by M/V Warren Time Event: 2024-05-06 16:06:00, Recovered by R/V Connecticut
Sensor Performance: From 2022-10-17 16:05:00 to 2024-05-06 16:06:00
Data_type: temperature and Depth: 50 m Data_type: conductivity and Depth: 50 m
Deployed on Mooring: A0144
Mooring Details: from 2021-06-17 02:45:00 to 2021-11-02 11:58:00
Time Event: 2021-06-17 02:45:00, Redeployed by R/V Connecticut Time Event: 2021-11-02 11:58:00, Recovered by M/V Warren Time Event: 2021-06-18 12:00:00, high battery voltage seen on buoy during daytime. probably solar panel connected to battery directly
Sensor Performance: From 2021-06-17 02:45:00 to 2021-11-02 11:58:00
Data_type: temperature and Depth: 10 m
Time:2021-06-17 02:45:00, Status: 0
Comments: instrument mistakenly deployed at actual depth of 6.3m (from pressure sensor and confirmed on recovery)
Data_type: conductivity and Depth: 10 m
Time:2021-06-17 02:45:00, Status: 0
Comments: instrument mistakenly deployed at actual depth of 6.3m (from pressure sensor and confirmed on recovery)
Non-Deployment Sensor Events:
Time Event: 2022-03-19 00:00:00, Event: In_from_Repair
Time Event: 2022-03-19 00:00:00, Event: In_from_Repair
Comments: Performed initial diagnostic evaluation. Performed pressure calibration. Performed "POST" cruise calibration. Replaced the O-rings. Performed a hydrostatic pressure test. Installed NEW AF24173 AFCs