2019-06-04 18:18:00 | ALL | 999 | ALL | deployed by the R/V Connecticut |
2019-06-10 21:00:00 | ALL | 999 | ALL | battery voltage high (>16.5V during deployment). Probably solar panel plugged directly into battery port |
2019-06-10 22:00:00 | ALL | 20 | sbe37 | Last data received from 20m SBE37 IM. |
2019-11-01 03:30:00 | ALL | 1 | sbe37 | Last data received from 1m SBE37 SM. |
2020-01-24 13:57:45 | ALL | 2 | aanderaa | last data from instrument, stopped reporting, clock was drifiting slowly, cause unknown. |
2020-05-16 00:00:00 | visibility | -3 | met | visibility response has decreased, very small response when very low (verified) visibility present (flagged 2020-06-29) rjf |
2020-11-10 12:50:00 | ALL | 999 | ALL | Recovered by R/V Connecticut |
2021-04-03 16:00:00 | ALL | 50 | sbe37 | Post-recovery processing completed for 50m SBE37IMP 14730. Low conductivity spikes persistent in last 4-5 months of deployment? |
2021-04-10 15:00:00 | ALL | 20 | sbe37 | Post-recovery processing completed from 20m SBEIM2568 - data in file ends one week after deployment (rest of samples in memory contain data from older deployments)... |
2021-04-10 15:00:00 | ALL | 1 | sbe37 | Post-recovery processing show no data in instrument from the deployment time period (only has data from previous deployments). Not correct S/N??? |